800MHz 54mm Narrow Bore
Frequency/Bore diameter/SelfShield 800MHz/54mm/No 800MHz/54mm/SS
Field Srength(T)
18.79
18.8
Field homogeneity(Hz/hr) <=15 <=10
Vertical 0.5mT Stray Field Contour(m) 8.46 4.29
Vertical 1mT Stray Field Contour(m) 6.72 -
Vertical 5mT Stray Field Contour(m) 3.95 -
Horizontal 0.5mT Stray Field Contour(m) 6.69 2.68
Horizontal 1mT Stray Field Contour(m) 5.30 -
Horizontal 5mT Stray Field Contour(m) 3.08 -
Minimum Nitrogen Re-fill Interval(day)/ Volume(L)/Evaporation rate(cc/hr)
19/237/520 24/297/500
Minimum Helium Re-fill Interval(day)/ Volume(L)/Evaporation rate(cc/hr)
60/108/75 50/160/120
SCM Height(mm) 3275 3230
Weight (Kg) 2800 5420
Cryogens Weight (Kg) 290 420
Minumum Operational Ceiling Height(mm) 4159 4150
Cold Shipping
No
-